Variation in the reporting of glucose values during simultaneous glucose sensor wear

7Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: We evaluated variations in glucose values recorded by simultaneously worn glucose sensors. Method: Children (n=44) randomised to the hybrid closed-loop arm of the CLOuD study (Closed-Loop from Onset of Diabetes) had multiple two-week periods of simultaneous sensor wear (sensors A and B or sensors C and D worn together) during unrestricted living. Glucose metrics were calculated from each sensor and paired sample T-tests were used to evaluate differences. Results: Sensor A recorded lower mean glucose by 0.4mmol/L (95% CI -0.8, -0.1) (mean±SD 7.6±0.5mmol/L vs 8.0±0.9mmol/L), 2.1 percentage points (95% CI -3.4, -0.9) less time <3.9mmol/L (median [IQR] 3.2% [2.3, 4.4] vs 4.3% [2.5, 7.4]), and 7.9 percentage points (95% CI 5.0,10.7) greater time 3.9–10.0mmol/L (78.7±5.1% vs 70.8±8.0%) than sensor B. Sensor C recorded higher mean glucose by 1.2mmol/L (95% CI 1.0, 1.5) (8.3±0.7mmol/L vs 7.1±2.7mmol/L), 9.9 percentage points (95% CI -11.8, -8.0) less time <3.9mmol/L (median [IQR] 2.2% [1.6, 3.4] vs 12.8% [8.0, 17.2]), and 3.9 percentage points (95% CI 2.2, 5.7) greater time 3.9–10.0mmol/L (73.1±6.9% vs 69.2±8.3%) than sensor D. Conclusions: There is significant variability in reporting of glucose values by different sensors worn during unrestricted living. Copyright © 2023 John Wiley & Sons.

Cite

CITATION STYLE

APA

Sutton, H., Boughton, C. K., Allen, J. M., Ware, J., Wilinska, M. E., Hartnell, S., … Hovorka, R. (2023). Variation in the reporting of glucose values during simultaneous glucose sensor wear. Practical Diabetes, 40(6), 12–16. https://doi.org/10.1002/pdi.2482

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free