Fresh gas formulae do not accurately predict end-tidal PCO2 in paediatric patients

9Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

Abstract

To determine the fresh gas flow (FGF) requirements in paediatric patients, we measured the FGFs needed to maintain distal end-tidal PCO2 (PetCO2) values at 30 and 38 mmHg in patients weighing between 3.8 and 20 kg ventilated with either a Sechrist Infant Ventilator IV-100B® or an Air-Shields Ventimeter® and a Mapleson D circuit. The FCF requirement was 500 ml·kg-1·min-1 to maintain a PetCO2 of 30 mmHg and 250 ml·kg-1·min-1 to maintain a PetCO2 of 38 mmHg when minute ventilation ≥ FGF. When these formulae were used in a subsequent group of similar patients, a wide variation in PetCO2 measurements were obtained. We conclude that the safest and most accurate approach to determine the FGF requirement of paediatric patients is to continuously monitor the PetCO2 in each patient and to adjust the FGF accordingly. © 1988 Canadian Anesthesiologists.

Cite

CITATION STYLE

APA

Badgwell, J. M., Wolf, A. R., McEvedy, B. A. B., Lerman, J., & Creighton, R. E. (1988). Fresh gas formulae do not accurately predict end-tidal PCO2 in paediatric patients. Canadian Journal of Anaesthesia, 35(6), 581–586. https://doi.org/10.1007/BF03020344