Comparison between measured tissue phantom ratio values and calculated from percent depth doses with and without peak scatter correction factor in a 6 MV beam

  • Narayanasamy G
  • Cruz W
  • Papanikolaou N
  • et al.
N/ACitations
Citations of this article
13Readers
Mendeley users who have this article in their library.

Abstract

The purpose of this study is to examine the accuracy of calculated tissue phantom ratio (TPR) data with measured TPR values of a 6MV photon beam. TPR was calculated from the measured percent depth dose (PDD) values using 2 methods – with and without correcting for the differences in peak scatter fraction (PSF). Mean error less than 1% was observed between the measured and calculated TPR values with the PSF correction, for all clinically relevant field sizes and depths. When not accounting for the PSF correction, mean difference between the measured and calculated TPR values was larger than 1% for square field sizes ranging from 3 cm to 10 cm.

Cite

CITATION STYLE

APA

Narayanasamy, G., Cruz, W., Papanikolaou, N., & Stathakis, S. (2015). Comparison between measured tissue phantom ratio values and calculated from percent depth doses with and without peak scatter correction factor in a 6 MV beam. International Journal of Cancer Therapy and Oncology, 3(2), 03024. https://doi.org/10.14319/ijcto.0302.4