Urea biosensor based on a CO2 microsensor

26Citations
Citations of this article
42Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Urea sensors based on electrodes in direct contact with the medium have limited long-term stability when exposed to complex media. Here, we present a urea biosensor based on urease immobilized in an alginate polymer, buffered at pH 6, and placed in front of a newly developed fast and sensitive CO2 microsensor, where the electrodes are shielded by a gas-permeable membrane. The CO2 produced by the urease in the presence of urea diffuses into the microsensor and is reduced at a Ag cathode. Oxygen interference is prevented by a Cr2+ trap. The 95% response time to changes in urea concentration was 120 s with a linear calibration curve in the range 0−1000 μM and a detection limit of 1 μM. The Ni2+ cofactor to improve sensor performance was continuously supplied from a reservoir behind the sensor tip. The stability of the urea sensor was optimized by the addition of bovine serum albumin as a stabilizer to the urease/alginate mixture that was cross-linked with glutaraldehyde and Ca2+ ions. This immobilization strategy resulted in about 70% of the initial urea sensor sensitivity after two weeks of continuous operation. The sensor was successfully tested in blood serum.

Cite

CITATION STYLE

APA

Fapyane, D., Berillo, D., Marty, J. L., & Revsbech, N. P. (2020). Urea biosensor based on a CO2 microsensor. ACS Omega, 5(42), 27582–27590. https://doi.org/10.1021/acsomega.0c04146

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free