Prediction of optimal atrioventricular delay in patients with implanted DDD pacemakers

88Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.

Abstract

In patients with an implanted DDD pacemaker (PM), the atrial contribution may be interrupted by too short an atrioventricular (AV) delay, and filling time may be shortened by too long an A V delay. The A V delay at which the end of the A wave on transmitral flow coincides with complete closure of the mitral valve may be optimal. The subjects were 15 patients [70.3 ± 12.3 (SD) years old] with an implanted DDD PM. Cardiac output (CO) and pulmonary capillary wedge pressure (PCWP) were measured by Swan-Ganz catheter. Transmitral flow was recorded by pulsed Doppler echocardiography. A V delay was prolonged stepwise by 25 msc. When the AV delay was set at 155 ± 26 ms, the end of the A wave coincided with complete closure of the mitral valve. When the A V delay was prolonged 25, 50, 75, and 100 ms from this A V delay, the interval between the end of the A wave and complete closure of mitral the valve was prolonged 16 ± 5, 39 ± 6, 65 ± 4 and 88 ± 5 ms, respectively (r = 0.97, P < 0.0001) and diastolic mitral regurgitation was observed during this period. Thus, the optimal AV delay may be predicted as follows: the slightly prolonged A V delay minus the interval between the end of the A wave and complete closure of the mitral valve. When the A V delay was set at 215 ms, there was a significant positive correlation between the predicted optimal AV delay (166 ± 23 ms) and the optimal AV delay (CO:161 ± 26 msec, r = 0.93, P < 0.0001, PCWP: 161 ± 28 msec, r = 0.95, P < 0.0001). In conclusion, optimal AV delay can be predicted by this simple formula: slightly prolonged A V delay minus the interval between end of A wave and complete closure of mitral valve at the A V delay setting.

Cite

CITATION STYLE

APA

Ishikawa, T., Sumita, S., Kimura, K., Kikuchi, M., Kosuge, M., Kuji, N., … Usui, T. (1999). Prediction of optimal atrioventricular delay in patients with implanted DDD pacemakers. PACE - Pacing and Clinical Electrophysiology, 22(9), 1365–1371. https://doi.org/10.1111/j.1540-8159.1999.tb00630.x

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free