Minimum volume of discard for valid blood sampling from indwelling arterial cannulae

21Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The volume of discard for valid blood sampling from indwelling femoral (leader cath) and radial (Venflon) arterial cannulae (with internal volume of cannula plus connecting tubing of approximately 2 ml) was investigated by analysing sequential 2-ml aliquots of 10-ml samples. The aliquots were analysed for pH, carbon dioxide tension, oxygen tension, standard bicarbonate, haemoglobin concentration, haemoglobin oxygen saturation and potassium concentration. Analysis of variance showed that, for these variables and these catheters, a valid blood sample was obtained after discarding 4 ml, but not after only 2 ml. The haemoglobin concentration, as measured by a Corning oximeter, provided good warning of inadequate discard. © 1987 British Journal of Anaesthesia.

Cite

CITATION STYLE

APA

Clapham, M. C. C., Willis, N., & Mapleson, W. W. (1987). Minimum volume of discard for valid blood sampling from indwelling arterial cannulae. British Journal of Anaesthesia, 59(2), 232–235. https://doi.org/10.1093/bja/59.2.232