Automation of a single-DNA molecule stretching device

8Citations
Citations of this article
39Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

We automate the manipulation of genomic-length DNA in a nanofluidic device based on real-time analysis of fluorescence images. In our protocol, individual molecules are picked from a microchannel and stretched with pN forces using pressure driven flows. The millimeter-long DNA fragments free flowing in micro- and nanofluidics emit low fluorescence and change shape, thus challenging the image analysis for machine vision. We demonstrate a set of image processing steps that increase the intrinsically low signal-to-noise ratio associated with single-molecule fluorescence microscopy. Furthermore, we demonstrate how to estimate the length of molecules by continuous real-time image stitching and how to increase the effective resolution of a pressure controller by pulse width modulation. The sequence of image-processing steps addresses the challenges of genomic-length DNA visualization; however, they should also be general to other applications of fluorescence-based microfluidics.

Cite

CITATION STYLE

APA

Sørensen, K. T., Lopacinska, J. M., Tommerup, N., Silahtaroglu, A., Kristensen, A., & Marie, R. (2015). Automation of a single-DNA molecule stretching device. Review of Scientific Instruments, 86(6). https://doi.org/10.1063/1.4922068