Abstract
We aimed to discuss and evaluate the technical success and efficacy of the ArtVentive endolumi-nal occlusion system (EOS) device for splenic embolization. A retrospective review was undertak-en for all patients in whom the EOS device was deployed for the purpose of splenic embolization. Data was collected by a search of splenic artery embolization procedures in the hospital comput-er database. Data was reviewed for all patients in whom an EOS plug was deployed. Patient de-mographics, technical aspects of the procedure and follow-up at one month were reviewed. We review the technical success and efficacy of this occlusion device. Six patients underwent splenic embolization with the EOS plug. There were 5 male and 1 female patients; age range was 24–88 years. Five 8 mm and one 5 mm EOS plugs were deployed for the occlusion of the splenic artery. The technical success rate was 100% occurring in all 6 splenic arteries. One patient underwent a second angiogram and subsequent splenectomy for persistent splenic hemorrhage. One patient had a subsequent splenectomy for bacteremia with the spleen as the suspected source. This ear-ly data supports the efficacy of the EOS plug for the embolization of the proximal splenic artery.
Cite
CITATION STYLE
Moriarty, H. K., Clements, W., Koukounaras, J., Goh, G. S., Joseph, T., & Phan, T. (2020, September 1). Artventive endoluminal occlusion system for proximal splenic artery embolization. Diagnostic and Interventional Radiology. AVES. https://doi.org/10.5152/dir.2020.19417