Rapid determination of parabens in water samples by ultra-high performance liquid chromatography coupled to time of flight mass spectrometry

13Citations
Citations of this article
30Readers
Mendeley users who have this article in their library.

Abstract

An analytical methodology has been developed and validated for the purpose of identifying and quantifying four parabens (methylparaben, ethylparaben. propylparaben and/i-butylparaben) in water samples. The combination of rotating disk sorptive extraction (RDSE) technology with ultra-high performance liquid chromatography (UHPLC), along with electrospray ionization source (ESI) and time of flight mass spectrometry (TOF/MS) in trap mode, allowed for eliminating derivatization processes and a reduction of the chromatographic time required, achieving a greener analytical method. In this method, detection limits and precision (%RSD) were lower than 0.018 (Ug Lr1 and lower than 9.7% for all the parabens, respectively, being better than similar works. Matrix effect and absolute recoveries were studied in tap and sewage water samples to observe suppressions of the signals for all analytes, and absolute recoveries were around 60%. This methodology was applied to the analysis of two sewage samples (punctual and composite) obtained from locations in Santiago, Chile.

Cite

CITATION STYLE

APA

Becerra-Herrera, M., Miranda, V., & Richter, P. (2020). Rapid determination of parabens in water samples by ultra-high performance liquid chromatography coupled to time of flight mass spectrometry. Analytical Sciences, 36(6), 675–679. https://doi.org/10.2116/ANALSCI.19P409

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free