Comparison of a near-infrared reflectance spectroscopy system and skin conductance measurements for in vivo estimation of skin hydration: a clinical study

  • Saknite I
  • Zavorins A
  • Zablocka I
  • et al.
N/ACitations
Citations of this article
11Readers
Mendeley users who have this article in their library.

Abstract

Diffuse reflectance spectroscopy system was developed for estimation of skin hydration in the near-infrared spectral range of 900-1700 nm. Experimental setup consisted of a near-infrared spectrometer, Y-type fiber optics probe with 1 detection and 6 illumination fibers, halogen-tungsten light source and a PC. By analyzing diffuse reflectance spectrum, a parameter representing skin hydration by performing baseline correction and calculating the area under the 1450 nm water absorption maximum is proposed. A clinical study was performed acquiring data of skin hydration of 39 patients' forearm skin. Results of the developed system are compared to results obtained by a commercial device based on skin conductance measurements.

Cite

CITATION STYLE

APA

Saknite, I., Zavorins, A., Zablocka, I., Kisis, J., & Spigulis, J. (2017). Comparison of a near-infrared reflectance spectroscopy system and skin conductance measurements for in vivo estimation of skin hydration: a clinical study. Journal of Biomedical Photonics & Engineering, 3(1), 010310. https://doi.org/10.18287/jbpe17.03.010310