Abstract
Introduction: To compare the measured dose distributions to calculated ones in dose-to-water (Dw) and dose-to-medium (Dm) reporting modes for simple plans and patient-specific intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) plans using ArcCHECK with a fixed phantom density. Methods: The recommended density value of 1.18 g/cm3 for Acuros XB and X-ray voxel Monte Carlo was assigned to ArcCHECK on CT images. A total of 45 simple plans, including a 1-field plan, a 3-field plan, a 4-field plan, a half-arc plan from 270˚ to 90˚, and a full-arc plan, were assessed. Subsequently, the patient-specific 96 IMRT and VMAT plans were evaluated. Gamma analysis with a 3% normalized global dose error and a 3 mm distance-to-agreement criteria (γ3%G/3mm) was performed in the Dw and Dm. The change in γ3%G/3mm between Dw and Dm were statistically analyzed using JMPPro11 software.
Cite
CITATION STYLE
Hirashima, H., Nakamura, M., Ishihara, Y., Mukumoto, N., Akimoto, M., Tomita, T., … Mizowaki, T. (2018). Comparison of Gamma Pass Rate between the Dose-to-Water and Dose-to-Medium Reporting Modes for Patient-Specific QA Using a Helical Diode Array Dosimeter with a Fixed Phantom Density. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 07(01), 74–86. https://doi.org/10.4236/ijmpcero.2018.71007