Abstract

Wearable electrocardiogram (ECG) is attracting much attention for monitoring heart diseases in healthcare and medical applications. However, an imbalance usually exists between the contact resistances of sensing electrodes, so that a common mode noise caused by external electromagnetic field can be converted into the ECG detection circuit as a differential mode interference voltage. In this study, after explaining the mechanism of how the common mode noise is converted to a differential mode interference voltage, the authors propose a circuit with cadmium sulphide photo-resistors for cancelling the imbalance between the contact resistances and confirm its validity by simulation experiment. As a result, the authors found that the interference voltage generated at the wearable ECG can be effectively reduced to a sufficient small level.

Cite

CITATION STYLE

APA

Noro, M., Anzai, D., & Wang, J. (2017). Common-mode noise cancellation circuit for wearable ECG. Healthcare Technology Letters, 4(2), 64–67. https://doi.org/10.1049/htl.2016.0083