Abstract
Objectives: To evaluate the effects on vascular enhancement of either a fixed rate (FR) or a fixed injection duration (FID) in single-pass (SP) contrast-enhanced abdominal multi-detector CT (CE-MDCT). Methods: Ninety-nine (54 M; 45 F; aged 18-86 years) patients with nontraumatic acute abdomen underwent a SP CE-MDCT after i.v. injection of 1.7 cc/Kg of a nonionic iodinated contrast media (370 mgI/mL) performed with either a FR (2 cc/s; Group A) or a FID (55 s; Group B). In both groups, patients were further stratified according to total body weight (kg) as follows: 40-60 (L); 61-80 (M); 81-100 (H). Signal- (SNR) and contrast-to-noise ratios (CNR) were calculated for the liver and for both abdominal aorta (AA) and main portal vein (MPV). Statistical analysis was performed by Student t- or Chi-square test for continuous and categorical data, respectively, whereas post hoc analysis was performed by the Mann-Whitney test (P < .05). Results: There were no significant differences in demographic and physical characteristics between Group A (n = 50; 53 ± 20 years; BMI = 23.4 ± 4.4) and Group B (n = 50; 51 ± 17 years; BMI 22.7 ± 4.2). Whereas overlapping findings were observed in the M sub-groups (n = 40), SNR and CNR were significantly higher (P
Author supplied keywords
Cite
CITATION STYLE
Camera, L., D’Ambrosio, V., Pinto, L., Paludi, A., Liuzzi, R., De Cicco, R., … Brunetti, A. (2025). Fixed rate vs fixed injection duration in single-pass contrast-enhanced abdominal multi-detector CT: effects on vascular enhancement. British Journal of Radiology, 98(1165), 114–123. https://doi.org/10.1093/bjr/tqae202