Abstract
We evaluated the effect of dextran on values for total protein in serum as measured by the biuret method with five widely used automated instruments: the American Monitor Parallel; the Du Pont aca II; the Roche Cobas-Bio; the Kodak Ektachem 400; and the Beckman Astra 8. Dextran concentrations as great as 25 or 30 g/L had relatively little or no influence on total protein measurements by the latter three instruments. Dextran concentrations exceeding 6 g/L caused falsely low results with the aca, whereas the Parallel gave falsely high results when the dextran concentration exceeded 2 g/L. The aca total protein procedure could be protected from the interference by dextran concentrations up to 30 g/L by injecting 0.4-0.8 mL of ethylene glycol directly into the reagent pack before sampling. However, we could not eliminate the interference with the Parallel procedure by any simple means; we thus recommend that it not be used for measuring total protein in serum samples from patients who are being treated with dextran.
Cite
CITATION STYLE
Barnes, D. B., Pierce, G. F., Lichti, D., Landt, M., Koenig, J., & Chan, K. M. (1985). Effects of dextran on five biuret-based procedures for total protein in serum. Clinical Chemistry, 31(12), 2018–2019. https://doi.org/10.1093/clinchem/31.12.2018
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.