Abstract
In this paper, we present a proof of concept study which demonstrates for the first time the possibility of recording magnetocardiography (MCG) signals with 4He vector optically pumped magnetometers (OPM) operated in a gradiometer mode. Resulting from a compromise between sensitivity, size and operability in a clinical environment, the developed magnetometers are based on the parametric resonance of helium in a zero magnetic field. Sensors are operated at room temperature and provide a tri-axis vector measurement of the magnetic field. Measured sensitivity is around 210 f T (√Hz)-1 in the bandwidth (2 Hz; 300 Hz). MCG signals from a phantom and two healthy subjects are successfully recorded. Human MCG data obtained with the OPMs are compared to reference electrocardiogram recordings: similar heart rates, shapes of the main patterns of the cardiac cycle (P/T waves, QRS complex) and QRS widths are obtained with both techniques.
Author supplied keywords
Cite
CITATION STYLE
Morales, S., Corsi, M. C., Fourcault, W., Bertrand, F., Cauffet, G., Gobbo, C., … Labyt, E. (2017). Magnetocardiography measurements with 4He vector optically pumped magnetometers at room temperature. Physics in Medicine and Biology, 62(18), 7267–7279. https://doi.org/10.1088/1361-6560/aa6459