Energy resolution improvement of CdTe detectors by using the principal component analysis technique

8Citations
Citations of this article
9Readers
Mendeley users who have this article in their library.
Get full text

Abstract

In this paper, we report on the application of the Principal Component Analysis (PCA) technique for the improvement of the γ-ray energy resolution of CdTe detectors. The PCA technique is used to estimate the amount of charge-trapping effect which is reflected in the shape of each detector pulse, thereby correcting for the charge-trapping effect. The details of the method are described and the results obtained with a CdTe detector are shown. We have achieved an energy resolution of 1.8 % (FWHM) at 662 keV with full detection efficiency from a 1 mm thick CdTe detector which gives an energy resolution of 4.5 % (FWHM) by using the standard pulse processing method.

Cite

CITATION STYLE

APA

Alharbi, T. (2018). Energy resolution improvement of CdTe detectors by using the principal component analysis technique. Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 882, 114–116. https://doi.org/10.1016/j.nima.2017.11.027