Monitoring intra-abdominal pressure (IAP) has become a standard in intensive care units. Correlation between the abdominal wall’s thickness (AWTh) and IAP has been reported previously. The abdominal wall can be modeled as a compound of parallel dielectric slabs; changes in their width have a direct effect on the reflection coefficient of the abdominal wall at microwave frequencies. This work describes the design of a reflectometry system and its proof-of-concept trial on five patients during laparoscopic surgery. The system complies with IEEE Std. C95.1-2005 concerning exposure of humans to microwave electromagnetic fields in controlled environments. The results putatively show an inverse correlation between IAP and the reflection coefficient, and a strong dependence on the body mass index. A better understanding of the dynamics in the changes of the AWTh (during intra-abdominal hypertension) will allow further development of a microwave-based technique for the continuous non-invasive indirect monitoring of IAP in critical patients.
CITATION STYLE
David, M., Raviv, A., Guttel, A., García Reyes, V., Simini, F., & Pracca, F. (2021). Non-invasive indirect monitoring of intra-abdominal pressure using microwave reflectometry: system design and proof-of-concept clinical trial. Journal of Clinical Monitoring and Computing, 35(6), 1437–1443. https://doi.org/10.1007/s10877-020-00605-9
Mendeley helps you to discover research relevant for your work.