Preliminary investigation of PAGAT polymer gel radionuclide dosimetry of Tc-99m

13Citations
Citations of this article
15Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

PAGAT polymer gel was investigated as a suitable dosimeter materials for measuring absorbed dose from the unsealed source radionuclide Tc-99m. Differing amounts of Tc-99m over the range of 25-5000 MBq were introduced into a normoxic polymer gel mixture (PAGAT) in sealed nitrogen-filled P6 glass vials. After irradiation the gels were evaluated using MRI more than 48 hours after preparation to allow for radioactive decay. The dose delivered to the vial was also calculated empirically. R2 versus total activity curves were obtained over a number of experiments and these were used to evaluate the relationship between the amount of gel polymerization and the dose deposited by the radionuclide. A linear response up to 1000 MBq (corresponding to 20Gy) was displayed and was still behaving monotonically at 5000 MBq. Polymer gels offer the potential to measure radiation dose three-dimensionally using MRI. © 2009 IOP Publishing Ltd.

Cite

CITATION STYLE

APA

Braun, K., Bailey, D., Hill, B., & Baldock, C. (2009). Preliminary investigation of PAGAT polymer gel radionuclide dosimetry of Tc-99m. Journal of Physics: Conference Series, 164. https://doi.org/10.1088/1742-6596/164/1/012050