Effect of windowing and zero-filled reconstruction of MRI data on spatial resolution and acquisition strategy

175Citations
Citations of this article
176Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Standard, MR spin-warp sampling strategies acquire data on a rectangular k-space grid. That method samples data from the "corners" of k-space, i.e., data that lie in a region of k-space outside of an ellipse just inscribed in the rectangular boundary. Illustrative calculations demonstrate that the data in the corners of k-space contribute to the useful resolution only if an interpolation method such as a zero-filled reconstruction is used. The consequences of this finding on data acquisition and data windowing strategies are discussed. A further implication of this result is that the spatial resolution of images reconstructed with zero-filling (but without radial windowing) is expected to display angular dependence, even when the phase- and frequency-encoded resolutions are identical. This hypothesis is experimentally verified with a slit geometry phantom. It is also observed that images reconstructed without zero-filling do not display the angular dependence of spatial resolution predicted solely by the maximal k-space extent of the raw data. The implications of these results for 3D contrastenhanced angiographic acquisitions with elliptical centric view ordering are explored with simulations. © 2001 Wiley-Liss, Inc.

Cite

CITATION STYLE

APA

Bernstein, M. A., Fain, S. B., & Riederer, S. J. (2001). Effect of windowing and zero-filled reconstruction of MRI data on spatial resolution and acquisition strategy. Journal of Magnetic Resonance Imaging, 14(3), 270–280. https://doi.org/10.1002/jmri.1183

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free