Measurement of dose-area product with GafChromic XR Type R film.

11Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Many of the newer X-ray machines are equipped with electronic means to provide dose-area product (DAP) information. For machines without that ability, an alternative method is to record radiation on a film that can handle a large amount of cumulative exposure. The use of GafChromic XR Type R film was investigated for this purpose by placing it at the X-ray tube assembly to record the radiation in interventional radiological procedures. Dose-area product was determined with a reflective densitometer and then with a flatbed scanner. Precisions were demonstrated to be 5% and 2%, respectively. In a comparison with the machine-recorded DAP, a regression analysis showed the validity of both techniques for values less than 1200 Gy-cm2.

Cite

CITATION STYLE

APA

Thomas, G., Li, Y., Chu, R. Y. L., Cheung, J. Y., Maqbool, F., Rabe, F., & Burns, G. S. (2005). Measurement of dose-area product with GafChromic XR Type R film. Journal of Applied Clinical Medical Physics / American College of Medical Physics, 6(3), 122–132. https://doi.org/10.1120/jacmp.v6i3.2047