Abstract
Liquid scintillation spectrometry is widely used for the analysis of alpha and beta emitting radionuclides. Robust calibration of liquid scintillation (LS) spectrometers is fundamental to accurate LS measurement but at the same time is time consuming and costly, particularly if a wide range of radionuclides are analysed by the laboratory. The frequency of the calibration varies in different laboratories and is based on many practical and operational factors. This work summarizes the observations regarding variations in 1220 Quantulus spectrometers efficiency calibrations performed annually using various radionuclides: 3H 63Ni, 55Fe, 36Cl, 45Ca, 147Pm, 241Pu, 99Tc for a period of 9 years and discusses the implication to calibration frequency.
Author supplied keywords
Cite
CITATION STYLE
Gaca, P., Warwick, P. E., & Croudace, I. W. (2017). Liquid scintillation counters calibration stability over long timescales. Journal of Radioanalytical and Nuclear Chemistry, 314(2), 753–760. https://doi.org/10.1007/s10967-017-5388-6