Real-Time Reflectance Measurement Using an Astigmatic Optical Profilometer

2Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

Abstract

An astigmatic optical profilometer with a commercial optical pickup head provides benefits, such as high resolution, compact size, and low cost. To eliminate artifacts caused by complex materials with different reflectances, a z-axis modulation mode is proposed to obtain quantitative surface morphology by measuring S curves on all image pixels. Moreover, the slope of the linear region in the S curve shows a positive relationship with the surface reflectance. However, the slope was calculated using an offline curve fitting method, which did not allow real-time reflectance imaging. Furthermore, quantitative reflectance data were unavailable because of the lack of calibration. In this study, we propose a novel method for real-time reflectance imaging by measuring the amplitude of a focus error signal (FES). The calibration results displayed a linear relationship between the FES amplitude and reflectance. The reflectance image of a grating sample with chrome patterns on a glass substrate demonstrates accurate reflectance measurements with a micrometer spatial resolution.

Cite

CITATION STYLE

APA

Liao, H. S., Huang, Y. K., Syu-Gu, J. Y., & Hwu, E. T. (2022). Real-Time Reflectance Measurement Using an Astigmatic Optical Profilometer. Sensors, 22(16). https://doi.org/10.3390/s22166242