Abstract
The accurate monitoring of phenylalanine concentration plays a prominent role in the treatment of phenylketonuria (PKU). In this study, we present an enzymatic assay based on Phenylalanine Dehydrogenase/NAD+ and tris (bipyridine) Ruthenium (II/III) as a colorimetric mediator for the detection of Phenylalanine concentration. The amount of amino acid was quantitatively recognized by optical absorption measurements at 452 nm through the conversion of Ru (byp)33+ to Ru (byp)32+, which is induced by the neoformed NADH. A detection limit of 0.33 µM, a limit of quantification of 1.01 µM, and a sensitivity of 36.6 a.u nM−1 were obtained. The proposed method was successfully tested using biological specimens from patients affected by hyperphenylalaninemia. The proposed enzymatic assay showed a high selectivity, making it a promising alternative for the development of versatile assays for the detection of phenylalanine in diluted serums.
Author supplied keywords
Cite
CITATION STYLE
Messina, M. A., Maugeri, L., Forte, G., Ruggieri, M., & Petralia, S. (2023). A highly sensitive colorimetric approach based on tris (bipyridine) Ruthenium (II/III) mediator for the enzymatic detection of phenylalanine. Frontiers in Chemistry, 11. https://doi.org/10.3389/fchem.2023.1164014