Abstract
C-reactive protein (CRP) is considered a marker of chronic inflammation and also a mediator of the atherosclerotic process. The purpose of this study is to compare the results of three quantitative immunoturbidimetric method used to measure CRP amounts in patient serum samples in a clinical laboratory. Freshly collected patient serums (n=100) were analyzed with 2 different analyzer (Siemens Advia 1800 and Abbott Architect C 8000) and 3 different reagents (Siemens, Sentinel, Archem). In order to determine whether they provide equivalent results in terms of traceability, the EP09-A3 guideline was used, and comparisons were evaluated within the scope of this standard. Limit of quantification (LoQ), Inter27 assay, Intra-assay, precision studies have been done. The relationship between these three reagents was determined by regression analysis and Bland Altman method. Regression coefficiencies between these three methods were found: ArchemSentinel r2=0,9987, Archem – Siemens r2=0,9986 and Sentinel – Siemens r2=0,9984. Regression equations between Archem-Sentinel y=-0,1359+1.0035x, between Archem-Siemens y=-0,02646+1,002x and between Sentinel – Siemens y=0,1326+0,9978x were found. For the first time in the literature, our results indicated that these three immunoturbidimetric methods were compatible.
Author supplied keywords
Cite
CITATION STYLE
Uca, E., Hacihasanoğlu Çakmak, N., & Yiğitbaşi, T. (2024). The comparison between the results of turbidimetric method for C-Reactive Protein measurement using different instruments. Acta Pharmaceutica Sciencia, 62(4), 778–792. https://doi.org/10.23893/1307-2080.aps6251