Abstract
Objectives: In this report, we evaluated utility of the capillary electrophoresis (CE) migration position of the CAPILLARYS 2 CE instrument. Methods: The precision of this x-axis number was determined on a selection of common hemoglobin (Hb) variants (Hb S, Hb C, Hb D-Punjab, Hb E, Hb Hope), and the reproducibility of this number was evaluated by comparing the results obtained by two large reference laboratories on 81 Hb variants. Additionally, the CE migration position is given for a total of 409 Hb variants. Results: The x-axis migration position showed excellent intra- and interassay precision. Comparison of Hb variants seen by both laboratories showed that 83% had a difference in migration position of 1 unit or less. Only three rare Hb variants showed a difference of more than 2 units. Conclusion: In summary, the CE migration position is a reproducible value and can be used as an aid in the identification of Hb variants.
Author supplied keywords
Cite
CITATION STYLE
Riou, J., Szuberski, J., Godart, C., Wajcman, H., Oliveira, J. L., Hoyer, J. D., & Bardakdjian-Michau, J. (2018). Precision of CAPILLARYS 2 for the detection of hemoglobin variants based on their migration positions. American Journal of Clinical Pathology, 149(2), 172–180. https://doi.org/10.1093/AJCP/AQX148