Validation of the noninvasive assessment of central blood pressure by the sphygmocor and omron devices against the invasive catheter measurement

83Citations
Citations of this article
80Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background To investigate the accuracy of the SphygmoCor and Omron HEM9000-AI devices in the estimation of central blood pressure (BP) in comparison with the simultaneous invasive catheter measurement. Methods The radial arterial pulse was sequentially recorded by the use of the Omron and SphygmoCor devices in 33 patients, with the calibration of the brachial oscillometric BP, to derive central BPs, which were also measured simultaneously with a catheter-based fluid-filled manometer system. The procedure was repeated three times to obtain 99 pairs of noninvasive and invasive measurements. Results The noninvasive central systolic BP estimations were significantly (P < 0.001) associated with the invasive measurement at the ascending aorta, with a correlation coefficient of 0.91 and 0.90 for the SphygmoCor and Omron devices, respectively. However, both devices underestimated central systolic BP with a difference of 15 mm Hg (95% confidence interval (CI), 17 to 13 mm Hg, P < 0.001) for SphygmoCor and 2 mm Hg (95% CI, 4 to 0 mm Hg, P < 0.05) for Omron. In comparison with the invasive catheter measurement at the brachial artery, the oscillometric Omron device underestimated brachial systolic BP by 19 mm Hg (95% CI, 23 to 15 mm Hg, P < 0.001). Conclusion sBoth devices underestimated central systolic BP, with a larger deviation by SphygmoCor. Nonetheless, these noninvasive estimations of central BP closely correlate with the invasive measurements, and can still be properly used, on the condition that device specific diagnostic thresholds become available. © 2011 American Journal of Hypertension, Ltd.

Cite

CITATION STYLE

APA

Ding, F. H., Fan, W. X., Zhang, R. Y., Zhang, Q., Li, Y., & Wang, J. G. (2011). Validation of the noninvasive assessment of central blood pressure by the sphygmocor and omron devices against the invasive catheter measurement. American Journal of Hypertension, 24(12), 1306–1311. https://doi.org/10.1038/ajh.2011.145

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free