Abstract
BACKGROUND: Many in vitro models report higher inhaled dose with dry versus heated humidity. Heat-and-moisture exchangers (HMEs) provide passive humidity in ventilator-dependent patients but act as a barrier to aerosol. The HMEs designed to allow aerosol delivery (HME-ADs) have not been well described. The purpose of this study is to determine the impact on aerosol deposition of HME-ADs with and without active exhaled humidity in a simulated ventilator-dependent adult model. METHODS: We used an in vitro lung model consisting of an intubated teaching mannequin with an endotracheal tube of 8.0 mm inner diameter with bronchi directly attached to a collecting filter and passive rubber test lung to provide testing without active exhaled humidity. To simulate exhaled humidity, a Cascade humidifier (37°C and 100% relative humidity) was placed between the collecting filter and test lung, simulating body temperature and pressure saturated exhaled humidity at the bronchi. Albuterol sulfate (2.5 mg/3 mL) was administered with a mesh nebulizer (Aerogen Solo) placed in the inspiratory limb of the ventilator circuit at the Y-piece, with no HME in place (control) and with 3 HME-AD devices, including the CircuVent, Humid-Flo, and AirLife, with and without exhaled humidity. Drug was eluted from the collecting filter and analyzed with spectrophotometry. Student t tests and analysis of variance were used for data analysis (P
Author supplied keywords
Cite
CITATION STYLE
Ari, A., Alwadeai, K. S., & Fink, J. B. (2017). Effects of heat and moisture exchangers and exhaled humidity on aerosol deposition in a simulated ventilator-dependent adult lung model. Respiratory Care, 62(5), 538–543. https://doi.org/10.4187/respcare.05015
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.