Abstract
Long term gamma spectroscopy experiments involving single-channel analyzer equipment depend upon thermal stability of the detector and its associated high-voltage supply. Assuming constant discrimination levels, a drift in the detector gain impacts the output rate, producing an effect on the output spectrum. In some cases (e.g. single-energy resonant absorption experiments) data of interest can be completely lost. We present a digital self-adapting discrimination strategy that tracks emission line shifts using statistical measurements on a predefined region-of-interest of the spectrum. It is developed in the form of a synthesizable module that can be intercalated in the digital processing chain. It requires a moderate to small amount of digital resources and can be easily activated and deactivated. © 2013 Springer Science+Business Media Dordrecht.
Author supplied keywords
Cite
CITATION STYLE
Veiga, A., Grunfeld, C. M., Pasquevich, G. A., Mendoza Zélis, P., Martínez, N., & Sánchez, F. H. (2014). Self-tuning digital Mössbauer detection system. Hyperfine Interactions, 224(1–3), 73–81. https://doi.org/10.1007/s10751-013-0771-3