Regional ventilation distribution in healthy lungs: can reference values be established for electrical impedance tomography parameters?

  • Yang L
  • Dai M
  • Cao X
  • et al.
N/ACitations
Citations of this article
29Readers
Mendeley users who have this article in their library.

Abstract

Background: Although electrical impedance tomography (EIT) is widely used for monitoring regional ventilation distribution, reference values have yet to be established for clinical use. The present study aimed to evaluate the feasibility of creating reference values for standard EIT parameters for potential clinical application. Methods: A total of 75 participants with healthy lungs were included in this prospective study (male:female, 48:27; age, 34±14 years; height, 172±7 cm; weight, 73±12 kg). The subjects were examined during spontaneous breathing in the supine position. EIT measurements were performed at the level of the 4 th intercostal space. Commonly used EIT-based parameters, including the center of ventilation (CoV), dorsal and most dorsal fractions of ventilation distribution (TV D and TV ROI4 respectively), global inhomogeneity (GI) index, and standard deviation of regional ventilation delay index (RVD SD) were calculated. Results: Following outlier detection, EIT data from 71 subjects were finally evaluated. The values of the evaluated parameters were: CoV, 48.7%±1.7%; TVD, 48.1%±5.4%; TV ROI4 , 7.1%±1.8%; GI, 0.49±0.04; and RVD SD , 7.0±2.0. The coefficients of variation for CoV and GI were low (0.03 and 0.07, respectively), but those for TV ROI4 and RVD SD were comparatively high (0.26 and 0.28, respectively). None of the evaluated parameters showed a significant correlation with age. The GI index showed a weak but significant correlation with body mass index (R=0.29, P=0.01). The RVD SD was slightly higher in males than in females. Conclusions: Our study indicated that CoV and GI were stable parameters with small coefficients of variation in participants with healthy lungs. The creation of EIT parameter reference values for setting treatment targets may be feasible.

Cite

CITATION STYLE

APA

Yang, L., Dai, M., Cao, X., Möller, K., Dargvainis, M., Frerichs, I., … Zhao, Z. (2021). Regional ventilation distribution in healthy lungs: can reference values be established for electrical impedance tomography parameters? Annals of Translational Medicine, 9(9), 789–789. https://doi.org/10.21037/atm-20-7442

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free