A simple device for monitoring neuromuscular blockade in children

2Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A simple instrument for measuring adductor pollicis tension in children has been devised. The thumb sensor utilizes a thin conical rubber seal obtained from a 5, 10 or 20 ml Becton Dickinson syringe. A syringe barrel of similar size, fixed at right angles to the sensor, forms a handle for the fist. The sensor is linked to a pressure transducer with noncompliant tubing. Pressure output of the salinefilled system was found to be linearly related to the force exerted on each size of the sensor. In 20 children, aged two to eight years, the device, fitted in turn with the 5,10 and 20 ml syringe seal sensors, gave train-of-four (TOF) ratios which showed close correlation (coefficient r = 0.96, 0.93, 0.95) and agreement (regression slope b = 0.97, 0.95, 0.98) with those of a force transducer. Thus, inspection of the displayed TOF pressure pulses gives a rapid and reliable estimate of neuromuscular blockade. © 1989 Canadian Anesthesiologists.

Cite

CITATION STYLE

APA

Lyew, M. A., Pinto, S. R., & Bevan, J. C. (1989). A simple device for monitoring neuromuscular blockade in children. Canadian Journal of Anaesthesia, 36(6), 717–721. https://doi.org/10.1007/BF03005429