Abstract
The importance and the difficulty of achieving good B0 homogeneity over the volume of interest in in vivo NMR spectroscopy are well known A fast, reliable, versatile, and fully automatic shimming procedure has been developed, using 3D chemical-shift imaging to measure the field distribution using only the water peak in the sample of interest, The procedure minimizes the mean-square error in the field distribution with respect to a constant held where the signal exists (e, g., the head). This produces a set of linear equations that can be solved by standard matrix methods. The procedure has been applied on a commercial imager producing water linewidths from the entire head as low as 8 Hz at 1.5 T. The mean linewidth and standard deviation from 94 head studies were 11.7 ± 1.9 Hz. © 1995 Academic Press, Inc.
Cite
CITATION STYLE
Hu, J., Javaid, T., Arias-Mendoza, F., Liu, Z., McNamara, R., & Brown, T. R. (1995). A Fast, Reliable, Automatic Shimming Procedure Using 1H Chemical-Shift-Imaging Spectroscopy. Journal of Magnetic Resonance, Series B, 108(3), 213–219. https://doi.org/10.1006/jmrb.1995.1126