Abstract
In a blind held evaluation of a nonisotopic liquid hybridization assay for detection of malaria parasites, 100 blood samples were tested from an area in which malaria is endemic; light microscopy was used as the reference test. Sensitivity, specificity, and positive and negative predictive values of the hybridization assay were 100%. One sample that was microscopy- negative and hybridization-positive was positive when reexamined. Another sample that was microscopy-positive and hybridization-negative was negative at reexamination. The detection limit of the test was ≤0.0005% parasitemia. Four samples with mixed infections were misdiagnosed by microscopy as single- species infections. Four samples diagnosed as mixed infections by microscopy and single infection by the hybridization test had no evidence of a second Plasmodium species upon reexamination. The polymerase chain reaction-based nonisotopic liquid hybridization assay was better than conventional light microscopy in detecting low-grade parasite infection and offers an exceptional advantage for detecting mixed infections.
Cite
CITATION STYLE
Oliveira, D. A., Shi, Y. P., Oloo, A. J., Boriga, D. A., Nahlen, B. L., Hawley, W. A., … Lal, A. A. (1996). Field evaluation of a polymerase chain reaction-based nonisotopic liquid hybridization assay for malaria diagnosis. Journal of Infectious Diseases, 173(5), 1284–1287. https://doi.org/10.1093/infdis/173.5.1284