Asymmetric Mach-Zehnder interferometer based biosensors for Aflatoxin M1 detection

54Citations
Citations of this article
74Readers
Mendeley users who have this article in their library.

Abstract

In this work, we present a study of Aflatoxin M1 detection by photonic biosensors based on Si3N4 Asymmetric Mach-Zehnder Interferometer (aMZI) functionalized with antibodies fragments (Fab'). We measured a best volumetric sensitivity of 104 rad/RIU, leading to a Limit of Detection below 5 × 10-7 RIU. On sensors functionalized with Fab', we performed specific and non-specific sensing measurements at various toxin concentrations. Reproducibility of the measurements and re-usability of the sensor were also investigated.

Cite

CITATION STYLE

APA

Chalyan, T., Guider, R., Pasquardini, L., Zanetti, M., Falke, F., Schreuder, E., … Pavesi, L. (2016). Asymmetric Mach-Zehnder interferometer based biosensors for Aflatoxin M1 detection. Biosensors, 6(1). https://doi.org/10.3390/bios6010001