Abstract
Introduction . Heavy metal intoxication after arthroplasty is extremely rare but could be lethal . Case Report . We report the case of a 69-year-old woman, who presented intense systemic symptoms of chromium-cobalt intoxication after revision of per-operative fractured ceramic components with metal-on-polyethylene. Systemic toxicity occurred a year after surgery and expressed brutally with mostly central neurological symptoms. Chelation associated with revision surgery allowed rapid regression of all symptoms. Conclusion . Revision of fractured ceramic, even per-operatively, should not be done with metal-on-polyethylene components, in order to avoid potentially lethal metal intoxication.
Cite
CITATION STYLE
Lecoanet, P., Blangis, M., Garcia, M., Legallois, Y., & Fabre, T. (2019). Chromium-Cobalt Intoxication with Intense Systemic Complications following Total Hip Revision after Per-Operative Ceramic Fracture. Case Reports in Orthopedics, 2019, 1–4. https://doi.org/10.1155/2019/4209796