In this contribution we present the wavelength calibration of the travelling reference Brewer spectrometer of the Regional Brewer Calibration Center for Europe (RBCC-E) at PTB in Braunschweig, Germany. The wavelength calibration is needed for the calculation of the ozone absorption coefficients used by the Brewer ozone algorithm. In order to validate the standard procedure for determining Brewer's wavelength scale, a calibration has been performed by using a tunable laser source at PTB in the framework of the EMRP project ENV59 ATMOZ Traceability for the total column ozone. Here we compare these results to those of the standard procedure for the wavelength calibration of the Brewer instrument. Such a comparison allows validating the standard methodology used for measuring the ozone absorption coefficient with respect to several assumptions. The results of the laser-based calibrations reproduces those obtained by the standard operational methodology and shows that there is an underestimation of 0.8ĝ€% of the ozone absorption coefficients due to the use of the parametrized slit functions.
CITATION STYLE
Redondas, A., Nevas, S., Berjón, A., Sildoja, M. M., Fabian León-Luis, S., Carreño, V., & Santana-Díaz, D. (2018). Wavelength calibration of Brewer spectrophotometer using a tunable pulsed laser and implications to the Brewer ozone retrieval. Atmospheric Measurement Techniques, 11(6), 3759–3768. https://doi.org/10.5194/amt-11-3759-2018