Investigate calibration methods for pressure transmitters

1Citations
Citations of this article
24Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Pressure transmitter[1-4] refers to a pressure sensor whose output is a standardized signal. The default memory value of the memory used to store the adjustment information in the internal signal conditioning chip is unknown, so the input and output curves of the pressure transmitter are unknown. The pressure transmitter must be calibrated to adjust the pressure transmitter input and output curves to standard input and output curves. Therefore, this paper studies a calibration method of the current output type pressure transmitter pressure based on the programmable gain amplifier PGA308. The method uses the average slope method to fit the pressure transmitter output output curve, which can realize the calibration of the pressure transmitter. The calibration accuracy can reach 0.0063%, and the calibration rate can reach 9.9 seconds/piece.

Cite

CITATION STYLE

APA

Wang, J., & Zhang, H. (2020). Investigate calibration methods for pressure transmitters. In Journal of Physics: Conference Series (Vol. 1646). IOP Publishing Ltd. https://doi.org/10.1088/1742-6596/1646/1/012084