Abstract
Background: Blood product safety was significantly improved by the introduction of NAT testing in the late 1990s, resulting in a strong decrease of transfusion-transmitted infections (TTIs). Due to the occurrence of HIV-1 NAT test failures as a consequence of mismatch mutations in the amplicon regions of mono-target NAT assays, the Paul Ehrlich Institute mandated the implementation of multi-target NAT assays for HIV-1 in 2014. Commercial suppliers mostly developed dual-target NAT assays, with only one implementing a triple-target NAT assay. Methods: The HIV-1 triple-target NAT assay v3 (GFE Blut) was tested on mutated specimens and synthetic DNA bearing mutations that resulted in sample underquantification or false-negative test results. In addition, data from 2 years routine testing at three German Red Cross Blood centres were analysed. Results: The HIV-1 triple-target PCR could compensate for all mutations tested and could compensate the loss of one amplicon without a significant loss of sensitivity. Data from 2 years routine testing showed a solid performance. Conclusion: The HIV-1 triple-target v3 assay (GFE Blut) can compensate mutations in target sequences better than a dual-target assay and is applicable to high-throughput screening, thus increasing blood product safety.
Author supplied keywords
Cite
CITATION STYLE
De Zolt, S., Thermann, R., Bangsow, T., Pichl, L., Müller, B., Jork, C., … Roth, W. K. (2016). Implementation of an HIV-1 Triple-Target NAT Assay in the Routine Screening at Three German Red Cross Blood Centres. Transfusion Medicine and Hemotherapy, 43(3), 183–189. https://doi.org/10.1159/000446290