Impact of new operational dosimetric quantities on individual monitoring services

9Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The new operational dosimetric quantities framework, proposed in the ICRU95 report, has stimulated the scientific community to start investigations that aim to assess its impact on radiation protection practices. As part of this effort, the present study describes an inter-comparison exercise among individual monitoring services (IMSs) on passive whole-body dosimetry. The inter-comparison is performed in terms of both the existing operational dose quantity Hp(10) and its proposed replacement Hp, to allow an evaluation of the actions that may be necessary to adapt dosimetry systems to the proposed quantity. For two of the tested IMSs, simple modifications to the detector response function, or the dose calculation algorithm, were sufficient to obtain results within acceptable limits. However, these approaches are not sufficient to give a level ofperformance comparable to that achieved in terms of Hp(10). This may require a modification to dosemeter design.

Cite

CITATION STYLE

APA

Caresana, M., Garlati, L., Zorloni, G., Behrens, R., Otto, T., Minchillo, G., & Rossi, F. (2021). Impact of new operational dosimetric quantities on individual monitoring services. Journal of Radiological Protection, 41(4). https://doi.org/10.1088/1361-6498/ac0d63