Localization of luminescent inhomogeneities in turbid media with spatially resolved measurements of cw diffuse luminescence emittance

  • Hull E
  • Nichols M
  • Foster T
62Citations
Citations of this article
23Readers
Mendeley users who have this article in their library.
Get full text

Abstract

We present a steady-state method for localizing a source ofluminescence (i.e., fluorescence or phosphorescence) buried in asemi-infinite turbid medium with unknown optical properties. Adiffusion theory expression describing the emittance of an isotropicpoint source is fit to spatially resolved surface measurements of thediffuse emittance from the luminescent source. The techniquereports the location of the center of a 6.0-mm-diameter, fluorophore-containing spherical bulb embedded in a liquid phantom withan accuracy of 1.0 mm or better for source depths as great as 40.0mm. Monte Carlo data are analyzed to investigate the range and thepossible sources of error in the reconstructed source depth.

Cite

CITATION STYLE

APA

Hull, E. L., Nichols, M. G., & Foster, T. H. (1998). Localization of luminescent inhomogeneities in turbid media with spatially resolved measurements of cw diffuse luminescence emittance. Applied Optics, 37(13), 2755. https://doi.org/10.1364/ao.37.002755

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley