Evaluation of a Novel Zonular Tension Restoring Accommodating Silicone IOL Design: Pilocarpine and Cyclopentolate-Induced Effect 20 Months after Implantation

3Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Purpose. To investigate a novel zonular-stress restoring accommodating 1-piece silicone IOL. Setting. Angeles City, Philippines. Design. Prospective randomized bilateral study. Methods. Each patient received a study IOL (ActaLens™, Emmetrope, La Canada, CA) in one eye and a control IOL (CrystaLens® AO, B&L, USA, or an AcrySof IQ®, Alcon, USA) in the contralateral eye to allow for intraindividual comparison. At the 20-month follow-up, two measurement days were set to measure all eyes before and after instilling 2% pilocarpine on the first day and 1% cyclopentolate on the second measurement day using an optical biometry device (Lenstar, Haag-Streit AG, Switzerland), respectively. PCO was graded by two examiners independently at the slit lamp. Results. In total, 16 eyes of 8 patients were included. In the study group and the control group, the pilocarpine-induced ACD shift was 0.32 mm (SD: 0.12) (p=0.014) and 0.04 mm (SD: 0.16) (p=0.854), respectively. In the study group and the control group, the mean cyclopentolate-induced ACD shift was 0.14 (SD: 0.06) (p=0.014) and 0.03 mm (SD: 0.03) (p=0.181), respectively. PCO and Nd: YAG rates were higher in the study group, but differences were not found to be significant (AcrySof vs. ActaLens p=0.100 and CrystaLens vs. ActaLens p=0.174). Conclusion. The investigated IOL is a novel concept for an accommodating IOL, and results showed a moderate pilocarpine-induced forward shift of the IOL 20 months following implantation. For all patients, the investigated IOL seems to have a higher PCO rate compared to standard monofocal IOLs.

Cite

CITATION STYLE

APA

Hirnschall, N., Henson, R. P. D., Vicencio, J. M. S., Angeles, A. L., Findl, O., & Phillips, A. (2021). Evaluation of a Novel Zonular Tension Restoring Accommodating Silicone IOL Design: Pilocarpine and Cyclopentolate-Induced Effect 20 Months after Implantation. Journal of Ophthalmology, 2021. https://doi.org/10.1155/2021/4797851

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free