Abstract
PURPOSE: To evaluate the elastic anisotropy of porcine corneas at different intraocular pressures (IOPs) using a noncontact optical coherence elastography (OCE) technique. METHODS: A focused air-pulse induced low amplitude (≤ 10 μm) elastic waves in fresh porcine corneas (n = 7) in situ in the whole eye globe configuration. A homebuilt phase-stabilized swept source optical coherence elastography (PhS-SSOCE) system imaged the elastic wave propagation at different stepped radial directions. A closed-loop feedback system was used to artificially control the IOP and the OCE measurements were repeated as the IOP was incrementally increased from 15 to 30 mm Hg in 5-mm Hg increments. RESULTS: The OCE measurements demonstrated that the stiffness of the cornea increased as a function of IOP and elastic anisotropy of the cornea became more pronounced at higher IOPs. The standard deviation of the modified planar anisotropy coefficient increased from 0.72 ± 0.42 at an IOP of 15 mm Hg to 1.58 ± 0.40 at 30 mm Hg. CONCLUSIONS: The presented noncontact OCE method was capable of detecting and assessing the corneal elastic anisotropy as a function of IOP. Due to the noninvasive nature and small amplitude of the elastic wave, this method may be able to provide further information about corneal health and integrity in vivo.
Cite
CITATION STYLE
Singh, M., Li, J., Han, Z., Wu, C., Aglyamov, S. R., Twa, M. D., & Larin, K. V. (2016). Investigating elastic anisotropy of the porcine cornea as a function of intraocular pressure with optical coherence elastography. Journal of Refractive Surgery, 32(8), 562–567. https://doi.org/10.3928/1081597X-20160520-01