Abstract
The aim of the present in vitro study was to assess the modes of failures under static load among titanium (Ti) and 1- and 2-piece zirconia abutments. The 1- and 2-piece zirconia abutment specimens were fabricated from prescanned Ti abutments. Twenty-one implant abutments and 21 implant replicas were equally divided into 3 groups as follows: (a) Group 1 (Titanium group); (b) Group 2 (1-piece zirconia abutment group); and (c) Group 3: 2-piece zirconia abutment group). A 250 000-cycle linear fatigue load ranging between 10 N and 210 N was applied to all specimens using an all-electric dynamic test instrument. The specimens were loaded until they fractured. In all groups, assessment of mode of fracture was done on visual assessment by a trained and calibrated investigator. Prior sample-size estimation was performed; and sample distribution was assessed using the Kolmogorov and Shapiro tests. Screw fracture (n ¼ 7) and abutment bending at the apical part (n ¼ 7) occurred in the Ti group. In the 1-piece zirconia group, screw and abutment fractures occurred in 7 and 7 cases, respectively. In the 2-piece zirconia screw fracture (n ¼ 7) above the Ti zirconia junction (transgingival segment) and abutment fracture (n ¼ 7) were determined as the failure modes. In vitro, the 1-piece zirconia abutments are more fracture resistant than titanium and 2-piece zirconia abutments. From a clinical perspective, further studies are needed to determine the minimum static load value required to induce fracture of the 1- and 2-piece zirconia abutments.
Author supplied keywords
Cite
CITATION STYLE
AlAmar, M., & Alqahtani, F. (2020). Effect of different implant-abutment connection materials on the fracture resistance of zirconia abutments. Journal of Oral Implantology, 46(2), 88–92. https://doi.org/10.1563/aaid-joi-D-19-00192
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.