Abstract
A new Field Imaging Spectrometer System (FISS) based on a cooling area CCD was developed. This paper describes the imaging principle, structural design, and main parameters of the FISS sensor. The FISS was spectrally calibrated with a double grating monochromator to determine the center wavelength and FWHM of each band. Calibration results showed that the spectral range of the FISS system is 437-902 nm, the number of channels is 344 and the spectral resolution of each channel is better than 5 nm. An integrating sphere was used to achieve absolute radiometric calibration of the FISS with less than 5% calibration error for each band. There are 215 channels with signal to noise ratios (SNRs) greater than 500 (62.5% of the bands). The results demonstrated that the FISS has achieved high performance that assures the feasibility of its practical use in various fields. © 2011 by the authors; licensee MDPI, Basel, Switzerland.
Author supplied keywords
Cite
CITATION STYLE
Zhang, L., Huang, C., Wu, T., Zhang, F., & Tong, Q. (2011). Laboratory calibration of a field imaging spectrometer system. Sensors, 11(3), 2408–2425. https://doi.org/10.3390/s110302408