Control system design for a continuous positive airway pressure ventilator

14Citations
Citations of this article
51Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Continuous Positive Airway Pressure (CPAP) ventilation remains a mainstay treatment for obstructive sleep apnea syndrome (OSAS). Good pressure stability and pressure reduction during exhalation are of major importance to ensure clinical efficacy and comfort of CPAP therapy. In this study an experimental CPAP ventilator was constructed using an application-specific CPAP blower/motor assembly and a microprocessor. To minimize pressure variations caused by spontaneous breathing as well as the uncomfortable feeling of exhaling against positive pressure, we developed a composite control approach including the feed forward compensator and feedback proportional-integral-derivative (PID) compensator to regulate the pressure delivered to OSAS patients. The Ziegler and Nichols method was used to tune PID controller parameters. And then we used a gas flow analyzer (VT PLUS HF) to test pressure curves, flow curves and pressure-volume loops for the proposed CPAP ventilator. The results showed that it met technical criteria for sleep apnea breathing therapy equipment. Finally, the study made a quantitative comparison of pressure stability between the experimental CPAP ventilator and commercially available CPAP devices. © 2012 Chen et al; licensee BioMed Central Ltd.

Cite

CITATION STYLE

APA

Chen, Z. L., Hu, Z. Y., & Dai, H. D. (2012). Control system design for a continuous positive airway pressure ventilator. BioMedical Engineering Online, 11. https://doi.org/10.1186/1475-925X-11-5

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free