Software-based compensation of computed tomography instrument misalignments – experimental study

  • et al.
N/ACitations
Citations of this article
6Readers
Mendeley users who have this article in their library.

Abstract

A newly developed software-based compensation of geometrical misalignments is applied to several computed tomography (CT) instruments. The actual geometry of each CT instrument is measured using a dedicated procedure. Then, given the measured geometries, software-based compensation is applied. Dimensional measurements on a validation object show a consistent reduction in observed errors as a result of the compensation. Robustness of the compensation method is confirmed also on a CT instrument with severe misalignments.

Cite

CITATION STYLE

APA

Ametova, E., Ferrucci, M., & Dewulf, W. (2019). Software-based compensation of computed tomography instrument misalignments – experimental study. E-Journal of Nondestructive Testing, 24(3). https://doi.org/10.58286/23723