Abstract
Infant warmer is one of the electromedical devices used to provide comfort and warmth to newborns, where the baby needs a temperature that is in accordance with the temperature in the mother's womb, which is between 34oC – 37oC. The purpose of this research is to study the Calibration technique and the calculation of the uncertainty of temperature measurement data.The results obtained from the calibration and calculations are ± 0,54oC for temperature. From these results, it was stated that the infant warmer was declared fit for use. As a rule, the test pass threshold value is ± 2 oC for temperature.
Cite
CITATION STYLE
Malau, M. S., & Yulizham, Y. (2021). ANALISA KALIBRASI ALAT INFANT WARMER MENGGUNAKAN ALAT INCUBATOR ANALYZER. JURNAL MUTIARA ELEKTROMEDIK, 5(2), 58–65. https://doi.org/10.51544/elektromedik.v5i2.3363