Design, manufacture and testing of capacitive pressure sensors for low-pressure measurement ranges

54Citations
Citations of this article
66Readers
Mendeley users who have this article in their library.

Abstract

This article presents the design, manufacture and testing of a capacitive pressure sensor with a high, tunable performance to low compressive loads (< 10 kPa) and a resolution of less than 0.5 kPa. Such a performance is required for the monitoring of treatment efficacy delivered by compression garments to treat or prevent medical conditions such as deep vein thrombosis, leg ulcers, varicose veins or hypertrophic scars. Current commercial sensors used in such medical applications have been found to be either impractical, costly or of insufficient resolution. A microstructured elastomer film of a polydimethylsiloxane (PDMS) blend with a tunable Young's modulus was used as the force-sensing dielectric medium. The resulting 18 mm × 18 mm parallel-plate capacitive pressure sensor was characterised in the range of 0.8 to 6.5 kPa. The microstructuring of the surface morphology of the elastomer film combined with the tuning of the Young's modulus of the PDMS blend is demonstrated to enhance the sensor performance achieving a 0.25 kPa pressure resolution and a 10 pF capacitive change under 6.5 kPa compressive load. The resulting sensor holds good potential for the targeted medical application.

Cite

CITATION STYLE

APA

Mitrakos, V., Macintyre, L., Denison, F. C., Hands, P. J. W., & Desmulliez, M. P. Y. (2017). Design, manufacture and testing of capacitive pressure sensors for low-pressure measurement ranges. Micromachines, 8(2). https://doi.org/10.3390/mi8020041

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free