Abstract
Solitaire FR device is a Food and Drug Administration-approved device for mechanical thrombectomy. It has been tested in various clinical trials for its safety and efficacy. We report a case of inadvertent detachment of the Solitaire FR device at stent-stent wire interface while performing mechanical thrombectomy. We review a rare phenomenon of retained Solitaire FR stent retriever in situ and discuss technique of avoidance and its management.
Cite
CITATION STYLE
Kinariwala, J., Rajah, G., & Luqman, A. (2018). Retained solitaire FR device after mechanical thrombectomy: Case review and management strategies. Brain Circulation, 4(4), 185. https://doi.org/10.4103/bc.bc_12_18