Abstract
Conventional single-voxel localization for MR spectroscopy (MRS) is restricted to selecting only rectangular-shaped regions of interest (ROIs). The complexity of tissue shapes of interest and the desire to maximize the signal-to-noise ratio (SNR) while minimizing partial-volume effects require more sophisticated localization techniques. A group of spatially selective RF pulses are proposed in this work for the measurement of spectra from regions of arbitrary shape based on using a radial trajectory in k-space. Utilizing a single k-line per excitation results in a broad spectroscopic bandwidth. However, spatial localization accuracy is compromised for nutation angles > 10° because of the small-tip-angle approximation of the Bloch equations. By interleaving multiple radial k-lines per excitation with nonselective refocusing pulses, one can achieve accurate localization for nutation angles up to 90° while simultaneously maintaining the spectral bandwidth. The technique is described and compared with existing localization methods, and in vivo results are demonstrated. © 2007 Wiley-Liss, Inc.
Author supplied keywords
Cite
CITATION STYLE
Qin, Q., Gore, J. C., Does, M. D., Avison, M. J., & De Graaf, R. A. (2007). 2D Arbitrary Shape-Selective Excitation Summed Spectroscopy (ASSESS). Magnetic Resonance in Medicine, 58(1), 19–26. https://doi.org/10.1002/mrm.21274