Successive gamma-ray irradiation and corresponding post-irradiation annealing of pMOS dosimeters

30Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

Abstract

The paper investigates a possibility of pMOS dosimeter re-use for the measurement of gamma-ray irradiation. The dosimeters were irradiated to the dose of 35 Gy, annealed at room and elevated temperatures, after which they were irradiated again to the same dose value. Changes in the threshold voltage shift during those processes were followed, and it was shown that their re-use depends on a gate polarization during irradiation. For the gate polarization of 5 V during irradiation the pMOS dosimeters can be re-used for measurements of the irradiation dose after annealing without prior calibration. The pMOS dosimeters with the gate polarization during irradiation of 2.5 V can also be re-used for irradiation dose measurements but they require calibration. It is shown that for their re-use it is necessary to anneal the pMOS dosimeter so that the fading is higher than 50%.